Industrial and medical dosimetry


Industrial dosimetry services

The High Dose Reference Laboratory (HDRL) services the radiation processing industry in Denmark and abroad, and in particular the medical device industry that sterilizes products using radiation.

HDRL is accredited by Danish Accreditation and Metrology Fund (DANAK) in accordance with EN ISO 17025:2017 for calibrations and dose measurements in the high-dose range 0.2 – 100 kGy.

Technical University of Denmark is appointed Designated Institute (DI) by the Danish Safety Technology Authority (“Sikkerhedsstyrelsen”) within 3 measurement areas: Length, Thermometry (Infrared) and Ionising Radiation (absorbed dose). HDRL maintains the latter within high-dose measurements for industrial applications.

Accreditation



SERVICES

Irradiations are carried out , at the:

  • Risø HDRL cobalt-60 gamma-cells,
  • 10 MeV electron accelerator or at the
  • Risø HDRL 100 keV electron accelerator

 

Risø HDRL gamma cells

Radiation source: Cobalt 60
The dose rates are traceable
to the national standards at
National Physical Laboratory, UK

 Gamma 1

Gamma 2 

Gamma 3

Source strength:

6.4 Gy/min (2014.01.01)

3.4 Gy/min (2014.01.01) 

164.8 Gy/min (2015.01.01)

Temperature control:

1: 10 - 80 oC
2: 50 - 200 oC

No

1: 10 – 80 oC

Maximum volume:

Height 18 cm
Diameter 12 cm
Height 15 cm
Diameter 12 cm

Height 20 cm
Diameter 15 cm

10 MeV electron accelerator

Risø HDRL collaborates with the industrial 10 MeV electron irradiation facility at Sterigenics, Denmark, for irradiation of dosimeters and samples.

Facility characteristics:    
Accelerator type: Rhodotron
Energy: 10 MeV
Beam current, average: max 8 mA
Maximum product size: 80 x 80 x 70 cm3

Sterigenics
Aa. Louis-Hansens Allé, 11,
3060, Espergaerde
Denmark
www.sterigenics.com    
     
Reference: ISO/ASTM 51649:2005 Practice for Dosimetry in an Electron-Beam Facility for Radiation Processing at Energies between 300 keV and 25 MeV

Risø HDRL low energy electron accelerator

The Risø HDRL low energy electron accelerator is used for irradiation of docimeters and samples.

Energy: 70-200 keV
Beam current: 1-10 mA
Conveyor tray speed: 3-30 m min-1
Maximum sample size: 20 x 30 cm2
Maximum sample height: 5 cm

References: ISO/ASTM 51818:2013 Standard Practice for Dosimetry in an Electron Beam Facility for Radiation Processing at Energies Between 80 and 300 keV

Guide on the Use of Low Energy Electron Beams for Microbiological Decontamination of Surfaces, Panel on Gamma and Electron Irradiation, 2013

 

Calibration of dosimeters by irradiation of the dosimeters at the user’s facility together with HDRL reference transfer dosimeters.

HDRL reference transfer dosimeters are:

Dichromate solution in ampoules

Made by: HDRL
Dose range: 10 - 50 kGy
Measurement: UV-Vis spectrophotometer
Pye-Unicam UV4
Use:  Reference dosimetry at gamma facilities
Reference:  ISO/ASTM 51401:2003 Practice for Use of a Dichromate Dosimetry System

Alanine pellets

Made by: Harwell dosimeters (UK)
Dose range: 0.2 - 100 kGy
Bruker EMS 104
Use: Reference dosimetry at gamma and electron facilities 

Alanine films

Made by:  Kodak 
Dose range:  1 - 120 kGy
Measurement: EPR spectrophotometer
Bruker EMS 104
Reference: ISO/ASTM 51607:2004 Practice for Use of an Alanine-EPR Dosimetry System

Calibrations are carried out in accordance with the recommendations given in the report CIRM 29 (Guidelines for the calibration of dosimeters for use in radiation processing).

Issue of calorimeters for routine or reference dose measurement at the user’s electron beam facility. HDRL supplies the software CalDose for dose measurement with calorimeters.

Calorimeters 

Made by: HRRL
Dose range:  Graphite: 1.5 - 15 kGy
Polystyrene: 3 - 40 kGy
Measurement: Digital ohm-meter
Use:  Routine and reference dosimetry at electron facilities
Reference: ISO/ASTM 51631:2003 Practice for Use of Calorimetric Dosimetry Systems for Electron Beam Dose Measurements and Routine Dosimeter Calibrations

Dosimetric validation of irradiation facilities for Installation and Operational Qualification (IQ – OQ) in accordance with the requirements in the international standard for radiation sterilization EN ISO 11137-1:2006.

Measurements of the fundamental characteristics and parameters of electron and gamma facilities.

References:

EN ISO 11137-1:2006
Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
ISO/ASTM 51649:2005
Practice for Dosimetry in an Electron-Beam Facility for Radiation Processing at Energies between 300 keV and 25 MeV. 
ISO/ASTM 51702:2004
Practice for Dosimetry in Gamma Irradiation Facilities for Radiation Processing
ISO/ASTM 51818:2002
Practice for Dosimetry in an Electron Beam Facility for Radiation Processing at Energies between 80 and 300 keV.

Guide on the Use of Low Energy Electron Beams for Microbiological Decontamination of Surfaces, Panel on Gamma and Electron Irradiation, 2013.

Measurement of dose distribution in products in order to ensure that all parts of the product are irradiated within the required specifications.

References:

EN ISO 11137-1:2006
Sterilization of health care products - Radiation - Part 1:
Requirements for development, validation and routine control of a sterilization process for medical devices 
ASTM E 2303-03  
Guide for Absorbed-Dose Mapping in Radiation Processing Facilities
Guide on the Use of Low Energy Electron Beams for Microbiological Decontamination of Surfaces, Panel on Gamma and Electron Irradiation, 2013.
Made by: Tesa, Hamburg 
Dose range: 10 - 100 kGy
Measurement: Scanner + RisøScan
Spectrophotometer
Use:  Measurement of dose distributions in irradiated product
Measurement of dose distributions in connection with IQ –
OQ at irradiation facilities
Reference:  ISO/ASTM 51275:2004. Practice for Use of a Radiochromic Film Dosimetry System

Software developed by Risø High Dose Reference Laboratory:

RisøScan 1.03 

RisøScan is a software package that is used for measurement of dose distributions based on analysis of images of visibly coloured dosimeter films, like FarWest, GAF and Risø B3 dosimeters. The images are created by scanning the dosimeter on a flatbed scanner.

RisøScan is used for dose mapping and for energy determination of high energy electron beams based on depth dose curves. The stability of the scanner is verified by use of a reference tablet.

For further information, please contact Risø High Dose Reference Laboratory.

References:

  • Helt-Hansen, J.; Miller, A., RisøScan - a new dosimetry software. 13. International Meeting on Radiation Processing, Chicago, IL (US), 7-12 Sep 2003. Radiat. Phys. Chem. (2004) 71 , 361-364
  • Helt-Hansen, J.; Miller, A., Assessment of dose measurement uncertainty using RisøScan. Radiat. Phys. Chem. (2006) 75, 1101-1106

CalDose 2.10 Beta

 
The Risø CalDose software is used to calculate the absorbed dose in Risø calorimeters based on resistance measurements before and after irradiation.

CalDose v.2.10 Beta is a beta-version. It is fully functional, except for the help-function. CalDose v.2.10 Beta is validated in accordance with the FDA guidelines (Title 21 CFR §820) - see the User Manual for details.

CalDose v.2.10 Beta is provided free of cost.

Download:

Right-click on the link below and select "save target as" to download the zip-file.
Extract the zip-file and follow the instructions in the User Manual on how to install CalDose.

 

Aluminium wedge for electron range and energy measurement at electron accelerator facilities. 

Reference: ISO/ASTM 51649:2005 Practice for Dosimetry in an Electron-Beam Facility for Radiation Processing at Energies between 300 keV and 25 MeV.

hdrl_other_dosimetry_equipment_jpg
Phantoms for dosimeter irradiation at electron accelerator facilities. 
hdrl_phantoms_elec_acc_jpg

Phantoms for dosimeter irradiation at gamma facilities.

Refence: CIRM 29 Guidelines for the calibration of dosimeters for use in radiation processing.

hdrl_phantoms_gamma_fac_jpg
Technical Note E1:
Stability of response of GEX DoseStix dosimeters heated after irradiation at different temperatures and different times. 
Technical Note E2:
Alanine dosimeters - Correction of response for irradiation temperature.
Technical Note E2.1
Alanine dosimeters - Correction of response for irradiation temperature.
Technical Note E3:
Evaluation of dose map for irradiation at verification dose.
Technical Note E4
Dose map for establishment of maximum acceptable dose.
Technical Note E5:

Dosimetry comparisons of Risø High Dose Reference Laboratory

Appendices for technical note E5

Appendix 1 - 2009 NPL - Risø irr e

Appendix 2 - 2010 NPL Risø irr e- (Panel)

Appendix 3 - 2012 NPL irr gamma - Risø

Appendix 4 - 2013 NPL - Risø irr gamma

Appendix 5 CCRI(I)-S2 Metrologia 48 (2011) Tech suppl 06009

Appendix 6 - 2014 NPL - Risø irr e-

Appendix 7 - 2015 NPL - Risø irr e- 

Appendix 8 - 2015b NPL irr gamma - Risø

Appendix 9 - 2016 NPL Risø irr e-

Appendix 10 - 2017 NPL - HDRL Risø irr gamma

 

Medical dosimetry calibration services

The Laboratory for Fundamental Medical Dosimetry (LFMD) within the DTU Dosimetry Reference Laboratory at DTU Health Tech offers DS/EN ISO/IEC 17025:2017 accredited calibrations of ionization chambers and electrometers in the context of radiotherapy dosimetry in accordance with the IAEA TRS-398 code of practice or similar protocols.

DTU Dosimetry Reference Laboratory is accredited by the Danish Accreditation and Metrology Fund (DANAK). The laboratory calibration and measurement capability (CMC) can be found here: DANAK database (accreditation reg. no. 266). The DANAK accreditation document can be found here (pdf file).

Ionization chamber calibrations are traceable to PTB in Germany.

Please send mail to indicate if you are potentially interested in calibrations at a given time.
For bookings, further information or discussions about the procedure, please contact: Claus E. Andersen.



SERVICES

DS/EN ISO/IEC 17025:2017 accredited calibration service

Ionization chamber (absorbed dose to water) in Co-60 beam.
Measurements of initial recombination and polarity are included in the certificate. 

Planned calibrations for ionization chambers is November/December and March/April. 

Please contact us to indicate if you are potentially interested in calibrations and at a given time (then will influence the planning of these activities).

For bookings and further information, please contact Claus E. Andersen

 

DS/EN ISO/IEC 17025:2017 accredited calibration service

Ionization chamber (absorbed dose to water) in Co-60 and five MV photon beams (4 MV - 18 MV). Measurements in Co-60 of initial recombination and polarity are included in the certificate.
Measurements in MV beams of volume recombination. 

Planned calibrations for ionization chambers is November/December and March/April. 

Please contact us to indicate if you are potentially interested in calibrations and at a given time (then will influence the planning of these activities).

For bookings and further information, please contact Claus E. Andersen

DS/EN ISO/IEC 17025:2017 accredited calibration service

Electrometer, charge calibration.

For bookings and further information, please contact Claus E. Andersen

DS/EN ISO/IEC 17025:2017 accredited calibration service

Electrometer, current calibration (DC). 

For bookings and further information, please contact Claus E. Andersen

Separate test of triax cables (leakage test for a range of high voltages, e.g. from 0 to +/- 500 V).

For bookings and further information, please contact Claus E. Andersen

Test of the full customer system (electrometer, ionization chamber and cable) in linac MV beams (and/or cobalt) against DTU reference measurements.

 

For example, it can be demonstrated if the current from the customer chamber under irradiation as measured with the customer electrometer system (after correction in accordance with the electrometer calibration certificate) at a given high-voltage setting is consistent with measurements performed with the same chamber and under the same conditions using the DTU master electrometer system and a DTU cable. The test could, for example, be in the 6 MV and/or 10 MV FFF (flattening-filter free) linac beams as to cover a range of clinically relevant dose-per-pulse conditions. Note that such system tests in cobalt are normally conducted under conditions of low (stable) currents (DC) whereas accelerator beams require that the electrometer can handle pulsed currents with high peak values. 

 

For bookings and further information, please contact Claus E. Andersen

Terabalt Cobalt-60 source

Cobalt-60 source for reference irradiations.
The cobalt irradiator is used as local reference. It has a well defined (almost monoenergetic) photon beam and a highly stable dose rate.

Model: Terabalt T100 from UJP Praha, Czech Republic.

Radionuclide: cobalt-60  (gamma energies: 1.17 MeV and 1.33 MeV)

Half-life of cobalt-60: 5.2710 y = 1925.21 d (u =0.015 % = 0.3 d) for 365.242 d/y.  This reference value is from Table of radionuclides (comments on evaluations) vol. 1-3, Monographie BIPM-5 (2006). The effective or "apparent" half-life describing the exponential decrease in dose rate at the reference point for the standardized DTU set-up is 1918.0 d (u=2.5 d with k=1).

Dose rate: ~1 Gy/min @ 100 cm distance (2018).

Beam: Fixed horizontal beam.

Collimator: 5x5 cm2to 40x40cm2.

Special features: Full computer control using the DTU scripting software (MELab).

A special rig has been designed for irradiations (see sketch below). The rig enables accurate positioning of ionization chamber free in air at the 1000.0 mm reference distance using the motorized xyz-stage. When the ionization chamber is in place, the water tank is lifted into position. 

DTU cobalt rig

Calibration of a PTW31021 ionization chamber in the cobalt beam.

DTU cobalt-60 irradiator 

The reference distance (1000.0 mm from the source) is realized using a micrometer.

Cobalt

Roos chambers and other plane parallel ionization chambers can be positioned at the iso-center line using a low-force indicator dial that previously has been set to zero using the optical telescope. The chamber can thereafter we moved 1.12 mm or 1.00 mm with the motorized xyz-stage in order to have the chamber at the desired reference point.

Cobalt

The details of the geometry for absorbed-dose to water calibrations are given below. The reference point is at 5.00 g/cm2of water. Distance C from the water tank front surface to the reference plane on the cobalt source head is established using a special tool involving an indicator dial with a well defined force.

DTU cobalt geometry

 

Varian Truebeam Linear Accelerator

Non-clinical medical linear accelerator for dosimetry research.
 

The Varian Truebeam is a state-of-the are medical linear accelerator. At Danish hospitals, similar accelerators are used for treatment of cancer patients (external beam radiotherapy). The Truebeam at DTU Health Tech is only for non-clinical work (i.e. no patients are treated here). We use the accelerator for dosimetry research and calibrations. 

The rig used for irradiation of ionization chambers and other dosimeters in MV photon beams. The Ionization chamber position can be moved using and a motorized xyz stage with sub-micron resolution, The control box and its joystick is shown to the right in the photo  above. The small rack to the right also contains a Keithley 2700 ohm meter for reading 8 thermistors used in the set-up plus a reference resistor.

Accelerator model: Truebeam from Varian Medical Systems.

Version: 2.0

Photon energies with flattening filter: 4, 6, 10, 15, and 18 MV

Photon energies without flattening filter (FFF): 6 and 10 MV

Electron energies: 6, 9, 12, 15, 16, 18, and 20 MeV

MLC: Millennium 120 leaf MLC (40 center pairs of 5 mm width, and 20 peripheral pairs of 10 mm width).

The accelerator iso-center is found before calibrations, and the distance from a reference plate is measured. The iso-center is transferred to the calibration rig using optical methods.

Calibrations are carried out with the water tank front surface 100.0 mm from the iso-center. This distance is realized as shown below using a calibrated 100.0 mm gauge block. One end of the block is positioned at the optical iso-center using an xyz motorized stage.The other end of the block is used for zeroing an indicator dial mounted on the accelerator head. When the water tank later is brought into position (see photos below), the water tank distance is then adjusted such that the indicatior once again shows zero. We therefore know that the water tank is at a 100.0 mm distance from the iso-center. A small adjustment can be applied, if a density-equivalent distance of 10 g/cm2is required. 

Linac

For TPR20.10measurements, we use a 200.0 mm gauge block. All linac calibrations are made relative to an external monitor chamber (PTW 7862) mounted in a special fixture on the accelerator head. The temperature of the monitor camber is measured using thermistors on both sides of the chamber.  

 Linac

The water tank surface has here been adjusted just that the indicator dial is back at zero.

Linac

Ionization chambers can be positioned with high precision at the iso-center line with or without sleeves. The two black manual xy-stages shown in the photo below contain cross-hair that have been adjusted to be on the iso-center line at the start of the calibration session. During the calibration session, these two cross-hairs facilitate verification that the Leica NA724 telescope remains on the optical iso-center.

Linac


Low-level electrical measurements system

Precision measurement of charge and current 
 

The key measurements in the medical dosimetry laboratory are carried out using ionization cambers connected to highly sensitive electrical measurement equipment.

We provide a calibration service for electrometer charge and current measurements..

DTU electrometer system

Three electrometers systems are used for calibrations in cobalt and MV photon beams from the linear accelerator: 

(1) The master electrometer is a Keithley 6517B electrometer with an external 1 nF low-leakage, air-filled, feedback capacitor. This air capacitor makes the system highly linear and stable as there is minimal influence of dielectric absorption normally found in other capacitors. Depending on the measurement task, the electrometer range can be set to 2 V, 20 V, or 200 V which corresponds to a maximum charge accumulation of 2 nC, 20 nC, or 200 nC, respectively. The system is calibrated in-situ against standards for DC voltage and capacitance, and the calibration therefore includes stray capacitance related to, for example, cables for the feedback capacitor. For DC current measurements, the charge accumulation is recorded every second triggered by an external 1 Hz frequency standard, and the linear regression slope of the charge vs. time recordings gives the current. With typical integration times from 30 s to 100 s, we can measure currents in the range from <10 fA to 6 nA. Current measurements in the femto ampere range are relevant mainly in the context of measuring cable and ionization chamber leakage.

A separate Keithley 6517 is used for the high voltage supply (up to 1000V), and the system can therefore be configured for floating or grounded measurements using connectors such as TNC, BNT, BNC/banana, or M-type. 

(2) A second electrometer system identical to the master system described above is used for independent checks.

(3) A third electrometer system based on a Keithley 6514 electrometer is used for recording the output from a PTW7862 external monitor chamber used during measurement in the linear accelerator. An important feature of this electrometer is that it covers the range up to 20µC. A separate Keithley  6517 is used for high-voltage supply.

The three electrometer systems are controlled using MELab software, a general-purpose data acquisition system developed by DTU Health Tech. For current measurements, the recording of charge vs. time takes place internally within each individual Keithley electrometer using so-called buffered data acquisition. The three electrometer systems are synchronized to the same external 1.000 Hz frequency standard.

Sleeves

Water-tight ionization chambers are calibrated directly in water unless otherwise requested.

High-voltage settings

Ionization chambers are calibrated at the main voltage requested by the customer.

Measurements with LFMD ionization chambers are normally carried out using a high-voltage setting such that negative charge is collected. For PTW30013, PTW31021, PTW34001, and FC65G ionization chambers we use 300 V, for NE2571 chambers we use 250 V, and for NPL2611 chambers we use 200 V.

Recombination measurements

Special procedures have been implemented for measuring the initial recombination in cobalt, and the initial and general recombination in linac beams (Niatel methid). These procedures involve measurement of the ionization chamber current at different dose rates. The initial recombination is estimated by extrapolation to zero dose rate.


Can we help you - please get in touch!

For information about services and prices please contact us.

Contact

Christina Ankjærgaard

Christina Ankjærgaard Head of Dosimetry, senior researcher Department of Health Technology Mobile: +45 61656929

Contact

Claus E. Andersen

Claus E. Andersen Senior researcher Department of Health Technology Phone: +45 46774912 Mobile: 2426 0311

Contact

Arne Miller

Arne Miller Specialist Consultant Department of Health Technology Phone: +45 46774224 Mobile: 2132 5613

Contact

Mark Bailey

Mark Bailey R&D Engineer Department of Health Technology Mobile: +459351 1480